The MS™ Graduated Stainless Steel Surgical Handle No.3
The Medi-Safe Scalpel handle is manufactured in accordance with the requirements of the ISO 7740 standard. (Instruments for surgery – Scalpels with detachable blades – Fitting dimensions)
The advanced automated manufacturing process assures that critical dimension tolerances and shape consistency are maintained throughout the manufactured batch.
Material: Stainless steel, grade 630 (17-4PH) Hardness: 21 Rockwell C minimum.
In accordance with the requirements of ISO 7153-1:1991
(Surgical Instruments – metallic materials Part 1– stainless steel)
Weight: 27-29 gram
The MS+ Scalpel Handle is designed and manufactured with the utmost care. It must be safely handled during and after surgery, including sterilization. To achieve long and satisfactory service, please ensure the instrument is only utilized as per its intended use.
The following instructions should be followed for best results.
Ultrasonic Cleaning:
Manual Cleaning:
Method 1:
For all instruments without other indications, steam autoclave sterilization with 2700F (1300C) for 15 minutes as recommended. Other temperatures and terms should be validated by the autoclave manufacturer. Sterilization temperatures should not exceed 2800F (1370C).
The instructions of the autoclave manufacturer must be observed also.
Method 2 :
Gamma Irradiation at 2.5 Mega Rad, follow validated process instructions.
Method 3:
Ethylene Oxide (ETO) Sterilization, follow validated process instructions.
NOTE
Contact the manufacturer of your Irradiator System to confirm appropriate sterilization parameters.
NOTE
Contact the manufacturer of your ETO sterilizer to confirm appropriate sterilization parameters